Understanding the Result Parameters
All result parameters are calculated from the BER threshold contour line and its
bounding box. So, all parameters change with the BER threshold.
The result display of the Eye Opening measurement shows the maximum eye
opening time, the maximum eye opening voltage, and the position of the optimum
sampling point.
These parameters are derived as illustrated in the figure below:
Error Location Capture
Error Location Capture - Concepts
The Error Location Capture measurement allows to capture the position of an
errored bit in a memory-pased pattern. The instrument saves the position of the
errored bit and writes a bit sequence neighbouring the errored bit to a file.
This feature can be used to find rare or random errors. A DUT could have problems
handling long series of zeroes. Error Location Capture can be used to locate the
bit errors in such cases.
6
Advanced Analysis
302
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...