This opens a dialog box which allows you to load a file. By default, the browser
searches for files with the suffix .jcs, but you can load any file.
N O T E
The required file format is described in detail in
“User-Defined Standards” on page
.
5
Check the
Show Jitter Curve
checkbox. It shows the jitter curve on the Jitter
Tolerance Characterization graph.
Before Running the Test
For DUTs that require a training sequence it is essential to prevent clock loss at
the beginning or end of the test. To avoid clock loss, the sinusoidal jitter component
has to be enabled. Since the sinusoidal jitter for the test is generated internally,
the jitter amplitude must be set to zero:
1
Click the
Jitter Setup
menu item from the
Jitter
submenu.
This opens the
Jitter Setup
window.
2
Enable jitter generation. Press the respective checkbox.
3
Activate
Sinusoidal
jitter.
4
Activate
Periodic
jitter
5
Set the jitter amplitude to 0 UI for both Sinusoidal jitter and Periodic jitter.
Running the Test
Once you have specified the
Frequency
,
BER
, and
Search
criteria, press the
Start
button to run the test.
The bar in the lower right-hand corner of the screen shows the progress.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
441
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...