How to Create the Calibration Settings
Use the following calibration settings:
• Preset Instrument State
• Enable De-Emphasis
• Set all cursors to 0dB
• Set Output Offset at N4916B => 0V
• Set Output Amplitude at N4916B => 400mV
• Set Data rate to the maximum possible data rate of your setup. This depends
on the maximum data rate of the PG, ED and N4916B. Note that a N4916B that
is specified for 10.5Gb/s should be calibrated at 12.5Gb/s (this makes selecting
the optimum input timing setting easier). A N4916B that is specified up to
14.2Gb/s shall bei either calibrated at the ED's maximum data rate (12.5Gb/s)
or the PG's maximum data rate (7Gb/s, 13.5Gb's or 14.2Gb/s).Use either the
PG's or ED's maximum data rate (whatever is less). If calibrating using an
oscilloscope, then use either 7Gb/s, 12.5Gb/s or 14.2Gb/s, depending on the
maximum data rate of the PG and the N4916B.
• Set Trigger Out: Pattern Trigger at Bit Position 0 (for the scope measurement)
• Set Trigger Out: Offset 0V
• Set Pattern: PRBS 2^7-1 for PG and ED
• Set Error Detector: Alignment BER Threshold 1E-1
• Set Error Detector: Nomal Input, Single Ended
• Set Error Detector: Pattern Sync Manual
Example Calibration Using a Sampling Oscilloscope
The following example shows the Input Timing Calibration for a target BER of
10
-3
and a confidence level of 95%. Lower target BER values will lead to more
reliable results for the settings that are close to the ‘worst’ Input Timing Setting.
But lower target BER values also lead to significantly longer measurement times.
Please refer to the notes at the end of the example for details about the required
measurement time at different target BER values.
The example is using the following instruments:
1
N4903B with option D14
2
N4916B specified for up to 14.2Gb/s
3
Infiniium DCA 86100A with a 83484A module
• Connect the N4916B with the N4903B using the N4915A-010 matched cable
pair
9
Solving Problems
484
Agilent J-BERT N4903B High-Performance Serial BERT
Setup Description
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...