Table 31
S.No
Parameter
Description
4
0-Level
Zero level is the
measurement of the mean
value of the logical 0 of an
eye diagram.
Note: This measurement is
made in a section of the
eye referred to, as the eye
window boundaries. The
default value for the NRZ
eye window is the central
20% of the bit period. You
can change the eye
window boundary settings
in the Configure
Measurement dialog box.
These settings determine
what portion of the eye will
be measured.
5
1-level
One Level is a measure of
the mean value of the
logical 1 of an eye diagram.
Note: This measurement is
made in a section of the
eye referred to, as the eye
window boundaries. The
default value for the NRZ
eye window is the central
20% of the bit period. You
can change the eye
window boundary settings
in the Configure
Measurement dialog box.
These settings determine
what portion of the eye will
be measured.
6
Eye Amplitude
The difference between
the logic 1 level and the
logic 0 level.
6
Advanced Analysis
338
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...