5
Corresponding to the M8061A entry, select the
Mux with Deemphasis
function
check box from the given options under
Enable Function
column. It opens the
M8061A Connection
dialog.
6
Click
Enable
button. It enables M8061A Multiplexer with Deemphasis that is
connected between data out of J-BERT N4903B and the DUT.
Once the
Mux with Deemphasis
function is enabled, you will see the
Mux with
Deemphasis
parameters in the M8061A window.
9
Solving Problems
520
Agilent J-BERT N4903B High-Performance Serial BERT