The parameters can be changed to configure the output according to the DUT’s
termination requirements. Be aware that the changes are not applied immediately
but only when the output protection circuit button is clicked.
Output Protection on M8061A Output(s)
The output protection state on M8061A output(s) occurs when the amplifier
detects an overload condition which is caused by the externally connected circuitry
(DUT). In the output protection state, the Auto Range, Amplitude Range, Amplitude,
High, Low, Offset, Coupling, Termination Model, Termination Voltage and CMI state
are re-programmed to safe values.
The GUI shows an error message, listing the effected outputs and highlights the
Outputs On
status indicator:
In the M8061A window, the parameters that currently do not match the data out
settings are highlighted in yellow.
The parameters can be changed to configure the output according to the DUT’s
termination requirements. The changes are applied only when the output
protection circuit button is clicked.
Setting up the Pattern Generator
4
Agilent J-BERT N4903B High-Performance Serial BERT
125
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...