Remote control commands
R&S
®
SMM100A
938
User Manual 1179.1341.02 ─ 05
Example:
SOUR:POW:ATT:RFOF:MODE FATT
uses maximum attenuation when the RF output is turned off.
Manual operation:
See
[:SOURce<hw>]:POWer:EMF:STATe
<State>
Displays the signal level as voltage of the EMF. The displayed value represents the
voltage over a 50 Ohm load.
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
n.a. (factory preset: 0)
Example:
POW:EMF:STAT 1
Activates voltage level display.
Manual operation:
See
"Display Level as Voltage of EMF"
[:SOURce<hw>]:POWer:LBEHaviour
<Behaviour>
Selects the level behavior at the RF output over time.
Parameters:
<Behaviour>
AUTO | UNINterrupted | MONotone | CVSWr | CPHase
UNINterrupted|MONotone
Do not use the uninterrupted level settings and strictly monotone
modes in combination with the high-quality optimization mode
(see
[:SOURce<hw>]:BB:IMPairment:OPTimization:
CVSWr
Constant VSWR
CPHase
Continuous phase
CONStant
Constant phase
*RST:
AUTO
Example:
SOURce1:POWer:LBEHaviour AUTO
Manual operation:
See
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
<Amplitude>
Limits the maximum RF output level in CW and sweep mode.
It does not influence the "Level" display or the response to the query
SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
.
SOURce subsystem