Multi-instrument setups
R&S
®
SMM100A
459
User Manual 1179.1341.02 ─ 05
To ensure that the remote connections to the external instruments will be automatically
established after instrument's start-up or a reboot, set
Remote command:
:SCONfiguration:EXTernal:REMote:CONNect[:ALL]
:SCONfiguration:EXTernal:REMote:DISConnect[:ALL]
Preset behavior: Keep connections to external instruments
Sets the effect of R&S
SMM100A preset ([Preset] or
*RST
) on the following:
●
The connection status between the instrument and the connected external instru-
ments.
●
The R&S
SMM100A itself and the preset values.
●
The connected instruments.
"Preset behavior..."
Off
On
R&S
SMM100A and preset
values
Standard R&S
SMM100A preset
.
Excluded from preset:
●
"I/Q Output Type"
Connection status
Terminated ("Disconnect All Remote")
Unchanged
Connected instruments
External instrument preset (
) External instrument preset
Remote command:
:SCONfiguration:EXTernal:PBEHaviour
Connector Name
Displays the connector name.
Remote command:
via keyword in the SCPI syntax, e.g.
IQOutput
Direction
Displays the connector direction as a color coded field.
Connector
Direction/Color
Enabled in the "I/Q Stream Mapping"
CODER
-
Input
Gray
-
I/Q
Yes
Not used
System configuration settings