Remote control commands
R&S
®
SMM100A
763
User Manual 1179.1341.02 ─ 05
// select the impedance for the external pulse modulation input/
// for the external pulse modulation trigger input
SOURce:PULM:SOURce EXT
SOURce:PULM:POLarity NORMal
SOURce:PULM:IMPedance G1K
SOURce:PULM:TRIGger:EXTernal:IMPedance G10K
// Configure the pulse generator settings
// Set pulse period, width, and delay
SOURce:PULM:PERiod 10 us
SOURce:PULM:WIDth 8 us
SOURce:PULM:DOUBle:WIDTh 0.0000012
SOURce:PULM:DOUBle:DELay 0.0000045
// Activate the signal output
SOURce:PGENerator:OUTPut:STATe 1
SOURce:PULM:STATe 1
OUTPut1:STATe 1
The following commands are available:
........................................................................................ 763
[:SOURce<hw>]:PULM:TRIGger:MODE
........................................................................... 764
....................................................................................... 764
........................................................................................ 764
........................................................................................ 764
[:SOURce<hw>]:PULM:DOUBle:DELay
............................................................................765
[:SOURce<hw>]:PULM:DOUBle:WIDTh
........................................................................... 765
[:SOURce<hw>]:PULM:DOUBle:STATe
............................................................................ 765
.........................................................................................765
..................................................................................... 766
........................................................................................ 766
[:SOURce<hw>]:PULM:OUTPut:VIDeo:POLarity
............................................................... 766
..................................................................................... 767
..................................................................................767
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance
...................................................... 767
[:SOURce<hw>]:PULM:MODE
<Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode>
SINGle | DOUBle
SINGle
Generates a single pulse.
DOUBle
Generates two pulses within one pulse period.
*RST:
SINGle
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
SOURce subsystem