Remote control commands
R&S
®
SMM100A
764
User Manual 1179.1341.02 ─ 05
[:SOURce<hw>]:PULM:TRIGger:MODE
<Mode>
Selects a trigger mode - auto, external, external single or external gated - for generat-
ing the modulation signal.
Parameters:
<Mode>
AUTO | EXTernal | EGATe | ESINgle
*RST:
AUTO
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:PERiod
<Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period>
float
The minimum value depends on the installed options R&S
SMM-
K22 or R&S
SMM-K23
Range:
20E-9 to 100
Increment: 5E-9
*RST:
10E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width>
float
Range:
20E-9 to 100
Increment: 10E-9
*RST:
2E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:DELay
<Delay>
Sets the pulse delay.
Parameters:
<Delay>
float
*RST:
1ms