Remote control commands
R&S
®
SMM100A
728
User Manual 1179.1341.02 ─ 05
:SCONfiguration:EXTernal:CODer<ch>:REMote:ICONnect
<InstrName>, <RfPath>
:SCONfiguration:EXTernal:FADer<ch>:REMote:ICONnect
<InstrName>, <RfPath>
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:ICONnect
<InstrName>,
<RfPath>
:SCONfiguration:EXTernal:RF:REMote:ICONnect
<InstrName>, <RfPath>
Selects an external instrument for the selected connector and triggers connection.
Parameters:
<InstrName>
string
Instrument alias name, as retrieved with the command
SCONfiguration:EXTernal:REMote:LIST?
the command
:SCONfiguration:EXTernal:REMote:ADD
.
<RfPath>
string
Determines the used RF output of the external instrument.
For a one-path instrument, the output path is fixed to RF1.
Example:
See
Example "Connecting and configuring external instruments"
Usage:
Setting only
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:FADer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:RF<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
Queries information on the external instrument.
Example:
SCONfiguration:EXTernal:BBMM1:REMote:INFO?
"SGT (101676)","LAN","rssgt100a101676","A",
"Rohde&Schwarz,SGT100A,1419.4501k02/101676,3.1.19.4-3.18.251.99"
Example:
See
Example "Connecting and configuring external instruments"
Usage:
Query only
Manual operation:
See
See
See
See
See
See
:SCONfiguration:EXTernal:CODer<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:FADer<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:RF:REMote:DISConnect
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:DISConnect
Disconnects the selected remote connection.
SCONfiguration subsystem