![Rohde & Schwarz SMM100A User Manual Download Page 461](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939461.webp)
Multi-instrument setups
R&S
®
SMM100A
461
User Manual 1179.1341.02 ─ 05
Remote Connection Status
Indicates the status of the remote connection. Click the status indicator to toggle the
remote connection status.
To connect/disconnect all external instruments, use the functions
.
Note:
External instruments connected to the R&S
SMM100A are preset (
)
whenever:
●
The connection to the external instrument is established or terminated ("Connect/
Disconnect All Remote" or "Rem. Conn > On/Off")
●
The R&S
SMM100A is preset ([Preset]).
Note:
An active "Rem. Conn." and an "RF State > On" do not mean that an I/Q stream
is mapped to this connector.
Use the "System Configuration > External Rf and I/Q > Display > Mapped Connectors"
filter to list only the connectors with a mapped I/Q stream.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:CODer<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:RCONnection:STATe?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:DISConnect
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
:SCONfiguration:EXTernal:RF:RCONnection:STATe?
:SCONfiguration:EXTernal:RF:REMote:DISConnect
:SCONfiguration:EXTernal:RF<ch>:REMote:INFO?
Instrument Name
Displays useful information on the connected external instrument, like the instrument
designation and the RF path the RF settings are coupled to.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:INAMe?
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:INAMe?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
:SCONfiguration:EXTernal:RF:INAMe?
:SCONfiguration:EXTernal:RF<ch>:REMote:INFO?
RF Couple
If enabled, this parameter couples all major RF setting (like the "RF Frequency", the
"RF Level" and the "RF State") of the external instrument to the R&S
SMM100A.
The R&S
SMM100A acts as a primary instrument, i.e the external instruments syn-
chronize their settings and the RF output state to the R&S
SMM100A. You can apply
delta frequency and level offset to the signal.
System configuration settings