Remote control commands
R&S
®
SMM100A
917
User Manual 1179.1341.02 ─ 05
// Deactivate the list mode
SOUR1:FREQ:MODE CW
.........................................................................................923
......................................................................................925
13.16.11.1
List mode settings
With the following commands, you can create list mode data, select the trigger mode
and determine the dwell time.
.......................................................................................... 917
................................................................................918
...................................................................................918
[:SOURce<hw>]:LIST:DWELl:LIST:POINts?
...................................................................... 918
...................................................................................919
[:SOURce<hw>]:LIST:FREQuency:POINts?
......................................................................919
............................................................................................919
[:SOURce<hw>]:LIST:INDex:STARt
................................................................................. 920
..................................................................................920
.........................................................................................920
.......................................................................................... 920
...........................................................................................921
.......................................................................................... 921
[:SOURce<hw>]:LIST:POWer:POINts?
.............................................................................921
[:SOURce<hw>]:LIST:TRIGger:EXECute
..........................................................................922
[:SOURce<hw>]:LIST:TRIGger:SOURce
.......................................................................... 922
..................................................................................... 923
[:SOURce<hw>]:LIST:DWELl
<Dwell>
Sets the global dwell time. The instrument generates the signal with the frequency /
power value pairs of each list entry for that particular period.
See also
Significant parameters and functions
.
Parameters:
<Dwell>
float
Range:
0.5E-3 to 100
Increment: 1E-6
*RST:
0.01
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
SOURce subsystem