Remote control commands
R&S
®
SMM100A
905
User Manual 1179.1341.02 ─ 05
Example:
LFO:BAND?
// BW10m
// the bandwidth of the externally supplied LF signal is 10 MHz
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:FREQuency
<Frequency>
Sets the frequency of the LF signal in
[:SOURce<hw>]:LFOutput:FREQuency:
CW|FIXed
mode.
Note:
●
If the LF generator is used as a signal source, the instrument performs the analog
modulations (AM/FM/
Φ
M/PM) with this frequency.
●
In sweep mode (
[:SOURce<hw>]:LFOutput:FREQuency:MODE
SWE
), the fre-
quency is coupled with the sweep frequency.
Parameters:
<Frequency>
float
Range:
0.1 to depends on the installed options
(R&S
SMM-K24)
Increment: 0.01
*RST:
1000
Default unit: Hz
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:PERiod?
Queries the repetition frequency of the sine signal.
Return values:
<LfSinePeriod>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
0.001
Default unit: s
Example:
See
Example "Configuring the LF generator"
Usage:
Query only
[:SOURce<hw>]:LFOutput:FREQuency:MANual
<Manual>
Sets the frequency of the subsequent sweep step if
Use a separate command for each sweep step.
SOURce subsystem