![Rohde & Schwarz SMM100A User Manual Download Page 340](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939340.webp)
Configuring the RF signal
R&S
®
SMM100A
340
User Manual 1179.1341.02 ─ 05
To determine the step size, select the parameter
"Linear"
Takes the frequency value entered as absolute value in Hz.
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
Dwell Time
Defines the duration of the individual sweep steps.
Note:
In case of considerable overrun conditions, the R&S
SMM100A turns off the
sweep mode.
See also
Chapter 7.10.4, "Significant parameters and functions"
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
RF level sweep:
[:SOURce<hw>]:SWEep:POWer:DWELl
LF frequency sweep:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl
Trigger Slope
For "Mode = Extern Step/Single", selects the polarity of the active slope of an applied
instrument trigger.
Trigger signal is expected at the Inst Trig connector.
"Positive"
The rising edge of the trigger signal triggers the instrument.
"Negative"
The falling edge of the trigger signal triggers the instrument.
Remote command:
[:SOURce<hw>]:INPut:TRIGger:SLOPe
Execute Single Sweep
In "Mode = Single", starts a sweep manually.
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
:TRIGger<hw>:FSWeep[:IMMediate]
RF level sweep:
[:SOURce<hw>]:SWEep:POWer:EXECute
:TRIGger<hw>:PSWeep[:IMMediate]
LF frequency sweep:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
:TRIGger<hw>:LFFSweep:IMMediate
General:
:TRIGger<hw>[:SWEep][:IMMediate]
Varying the RF signal in list or sweep mode