Configuring the internal baseband source
R&S
®
SMM100A
225
User Manual 1179.1341.02 ─ 05
4. A segment ("Current Segment") is output continuously as long as the value of the
parameter "Segment" is not changed. Change the value of the parameter "ARB >
Trigger In (MSW) > Segment" to select the segment index processed after com-
pleting the current one.
5. The seamless transition between the segments requires waveforms with equal
clock rate.
If the waveforms feature different sample rates, select "ARB > Multi Segment >
Level/Clock/Marker > Clock Mode = Highest" or "Clock Mode = User" to resample
the waveforms.
To enable an external triggering for high-speed switch over
To test DUTs/chips using different test signals at high throughput requires fast switch
over (for instance when testing ATE devices during manufacture).
We assume, that the required multi-segment sequence is created and loaded in the
ARB (see
"Generating a multi-segment waveform file (basic workflow)"
The segments are
output in incremental order
. The instrument is operated manually,
the ARB generator is in active state, and the instrument output is switched on. An
external global or local trigger source is connected to the instrument and the corre-
sponding connectors are configured.
The following settings enable
switching times of approx. 5 us
and the test signals
are scrolled through with the aid of an external trigger signal:
1. Select "ARB > Trigger In (MSW) > Next Segment Mode = Next Segment".
2. Select "ARB > Trigger In (MSW) > Next Segment Source = External Global Next
Segment 1".
3. Do not change the displayed segment index (parameter "ARB > Trigger In (MSW)
> Segment").
4. Enable "ARB > Trigger > Trigger Mode = Auto"
5. Select "ARB > Trigger > Trigger Source = External Global/Local Trigger"
6. If the waveforms in the multi-segment file feature different sample rates, select
"ARB > Multi Segment > Level/Clock/Marker > Clock Mode = Highest" or "Clock
Mode = User" to resample the waveforms.
The switch over between the segments is triggered by receiving of an external trig-
ger event. The segments are output in incremental order, i.e. Seg#0, Seg#1, etc.
The currently processed segment ("Current Segment") is output continuously as
long as a new trigger event is received. A new trigger event aborts the output of the
current segment immediately and starts the processing of the next one in the
sequence.
Tip:
To output the segments in any order, use the ARB sequencer mode, see
configure a multi-segment sequence using the play list function"
For general description on the configuration of the trigger and connector settings:
See also
Chapter 11.2, "Configuring local and global connectors"
Generating multi-segment waveform files