Remote control commands
R&S
®
SMM100A
922
User Manual 1179.1341.02 ─ 05
Return values:
<Points>
integer
Range:
0 to INT_MAX
*RST:
0
Example:
See
Example "Create an RF list and activate the list mode"
Usage:
Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See
Example "Create an RF list and activate the list mode"
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:LIST:TRIGger:SOURce
<Source>
Selects the trigger source for processing lists.
The designation of the parameters correspond to those in sweep mode. SCPI standard
uses other designations for the parameters, which are also accepted by the instrument.
The SCPI designation should be used if compatibility is an important consideration. For
an overview, see the following table:
Rohde
&
Schwarz parameter
SCPI parameter
Applies to the list mode parameters:
AUTO
IMMediate
[:SOURce<hw>]:LIST:MODE AUTO
SINGle
BUS
[:SOURce<hw>]:LIST:MODE AUTO
or
[:SOURce<hw>]:LIST:MODE STEP
EXTernal
EXTernal
[:SOURce<hw>]:LIST:MODE AUTO
or
[:SOURce<hw>]:LIST:MODE STEP
Parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list is restarted as soon as it is fin-
ished.
SINGle|BUS
The list is triggered by the command
EXTernal
The list is triggered externally and executed once.
*RST:
AUTO