![Rohde & Schwarz SMM100A User Manual Download Page 244](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939244.webp)
Configuring the internal baseband source
R&S
®
SMM100A
244
User Manual 1179.1341.02 ─ 05
Phase Step
Sets the step width that is used to increment the phase.
The resulting phase in the carrier table equals:
Phase =
"Phase Start" + n*"Phase Step"
n
is a value between 0 and the difference of "Carrier Stop" and "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
Delay Start
Sets the delay of the carrier marked by "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay[:STARt]
Delay Step
Sets the step width that is used to increment the delay.
The resulting delay in the carrier table equals:
Delay =
"Delay Start" + n*"Delay Step"
,
n
is a value between 0 and the difference of "Carrier Stop" and "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay:STEP
Input Waveform File
Accesses the standard "File Select" function for selecting the input file. The input file
contains the I/Q signal to be modulated onto
all carriers
of the selected carrier range.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:FILE
Apply Assistant Settings
Transfers the assistant settings to the carrier table.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:EXECute
4.10.3.4
Carrier graph
Access:
1. Select "Baseband > ARB > General > Multi Carrier".
Generating multi-carrier signals