Remote control commands
R&S
®
SMM100A
737
User Manual 1179.1341.02 ─ 05
[:SOURce]:INPut:USER:CLOCk:LEVel
<Level>
[:SOURce]:INPut:USER:TRIGger:LEVel
<Level>
Sets the threshold for any input signal at the User1-3 connectors.
Parameters:
<Level>
float
Range:
0.1 to 2
Increment: 0.1
*RST:
1
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
[:SOURce]:INPut:USER:PULM:LEVel
<Level>
Parameters:
<Level>
float
Range:
0.1 to 2
Increment: 0.1
*RST:
1
Default unit: V
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
"Threshold User4-6 Input/Threshold Pulse Input"
[:SOURce]:INPut:USER:CLOCk:IMPedance
<Impedance>
[:SOURce]:INPut:USER:TRIGger:IMPedance
<Impedance>
Selects the input impedance for the external trigger inputs.
Parameters:
<Impedance>
G1K | G50
*RST:
G1K
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
"Impedance Clock/Trigger Input"
[:SOURce]:INPut:USER:CLOCk:SLOPe
<Slope>
Sets the polarity of the active slope of an externally applied clock signal.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
SOURce subsystem