Configuring the RF signal
R&S
®
SMM100A
403
User Manual 1179.1341.02 ─ 05
No signal is applied to the sensor during zeroing. RF output is temporarily switched
off during that time.
When unchecked, the zeroing procedure is skipped. However, the RF signal level
can be blanked shortly. This setting is useful if blanking of RF is undesirable or the
absence of power at the sensor cannot be guaranteed.
●
"Execute"
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
Remote command:
[:SOURce<hw>]:CORRection:ZERoing:STATe
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Used SMM100A Settings For Measurement
Displays the settings relevant for the measurement.
"Modulation"
Indicates the modulation state
"Amplitude"
Displays the currently set level.
Remote command:
"Offset"
Displays the level offset, set with
Remote command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
"UCOR Refer-
ence Level"
Displays the reference value for user correction.
The user correction reference level is the set
7.12.3.4
Import/export list files
Access:
1. Select one of the following:
● "RF" > "Sweep/List" > "List mode".
● "RF" > "RF Level" > "User Correction".
Improving level performance