Network operation and remote control
R&S
®
SMM100A
652
User Manual 1179.1341.02 ─ 05
#NO_COMMAND_CODE_START
//no SCPI command available for parameter %PARAMETER !
#NO_COMMAND_CODE_END
#EXIT_CODE_START
viClose (handle);
viClose (defaultRM);
return 0;
}
#EXIT_CODE_END
12.11.3
Remote control states
How to recognize if there is an active remote connection to the instrument
► Observe the indication on the taskbar.
A softkey in the taskbar indicates if and what kind of remote connections are cur-
rently set up.
See also
Chapter 12.5.6, "Remote connections settings"
The following table shows the different remote control states and the associated com-
mands or actions to return to manual control.
Table 12-8: Remote control status icons
GUI symbol
Transition to remote control state...
Transition to manual operation...(local state)
>R
(controller)
Remote control, but usable front panel keys.
The parameters are in read-only mode.
●
>L
(controller)
●
Tap the "Remote" icon (display)
●
[Local] (front panel or key emulation)
A currently performed setting is indicated by the green
arrows.
The setting must be completed (white arrows), other-
wise the instrument remains in remote state.
&LLO
(controller)
Remote control with locked front panel keys to
prevent user interaction.
The parameters are in read-only mode.
You can unlock LLO, and thus return to manual
operation only via remote control.
●
&LOCS
(controller)
●
CALL IBLOC (generator%)
(controller)
Note:
The command
&REMS
returns to "Remote" state.
References