Configuring the internal baseband source
R&S
®
SMM100A
98
User Manual 1179.1341.02 ─ 05
Figure 4-6: Trigger mode retrigger
* = Internal trigger event (to simplify the description)
Compare the shape and the length of the red curve in single mode and in retrigger
mode. The first part of the curve in retrigger mode is shorter. The signal generation
is interrupted due to the subsequent trigger event.
●
Armed auto
The instrument starts the signal generation only when a trigger event occurs and
then generates a continuous signal. An "Arm" stops signal generation. A subse-
quent trigger event causes a restart of the signal generation.
Figure 4-7: Trigger mode armed auto
* = Internal trigger event (to simplify the description)
●
Armed retrigger
The instrument starts the signal generation only when a trigger event occurs and
then generates a continuous signal. Every subsequent trigger event causes a
restart of the signal generation.
An "Arm" stops the signal generation. A subsequent trigger event causes a restart.