Configuring the RF signal
R&S
®
SMM100A
401
User Manual 1179.1341.02 ─ 05
Edit ← Data handling keys
Enables you to insert, or delete a row or ranges within a list, and provides access to a
dialog for automatic filling, see
Fill with Sensor ← Data handling keys
Opens a dialog to configure the automatic filling of user correction data with an R&S
NRP power sensor. Available in UCOR mode only.
See
Chapter 7.12.3.3, "Fill with sensor"
Save As/Save ← Data handling keys
Stores the list in a file with user-defined name and predefined file extension. To save a
copy or create a file, use the "Save as" function.
Fill...
Provides parameters for filling a table automatically with user-defined values.
To fill the table, select "Fill".
Note:
Once you enter a value or fill a column, the editor automatically adds preset val-
ues in the other columns. This functionality protects against data loss, otherwise
incomplete rows will be lost when saving. You can simply override these values.
"From / Range"
Defines the start line and number of lines to be filled.
"Column To Fill"
Selects, if the column is filled up with frequencies in Hz, levels in dBm
or dwell times in s.
"Start Value / End Value"
Sets the start value for frequency, level or dwell time. The end value
is read only and depends on the increment value and the range.
"Increment Value"
Determines the step size.
"Fill"
Fills the column specified in "Column To fill".
Improving level performance