File and data management
R&S
®
SMM100A
488
User Manual 1179.1341.02 ─ 05
► To reset an individual parameter: Open its context-sensitive menu and select "Pre-
set this parameter...".
●
"Factory Preset"
A factory preset is the most profound preset function that resets almost all instru-
ment settings, including reference oscillator, network and remote access settings.
Retained are the following settings:
–
Security, password, and settings protected by these passwords
–
User-defined data, like setups or data lists
–
Settings that relate to an integration of the instrument in a measurement setup.
► To restore the factory defaults, select
System Config > Setup > Settings > Factory
Note:
Perform a "Factory Preset" only if it is necessary. After a "Factory Preset",
the network connection to the instrument no longer exists.
Presetting the instrument to a user-defined instrument state
The reset functions set the parameters and operating modes to default values prede-
fined by the factory. Alternatively to these default settings, you can:
●
Define user-specific recall settings to be restored after a preset
(see
Chapter 10.2.3, "How to recall user settings automatically after preset"
●
Store and reload user-defined instrument states
(see
Chapter 10.4.2, "How to save and recall instrument settings"
Mark / Do not mark parameters changed from preset
To survey the current state of the settings concerning default values, the
R&S
SMM100A offers a feature that visually identifies deviations from the default val-
ues.
For more information, see
Chapter 10.2.2, "How to identify parameters which are not in
10.2.1
Preset, set to default and factory preset settings
...............................................................................................489
Preset
Resets all parameters and switching states, and closes all opened dialogs.
Restoring the (default) instrument configuration