Remote control commands
R&S
®
SMM100A
738
User Manual 1179.1341.02 ─ 05
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
[:SOURce]:INPut:USER:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Example:
See
Example "Global connectors settings configuration"
Manual operation:
See
[:SOURce<hw>]:INPut:TM:CLOCk:LEVel
<Level>
[:SOURce<hw>]:INPut:TM:TRIGger:LEVel
<Level>
Sets the high/low threshold in volts for the trigger and clock signal inputs of the base-
band section.
Suffix:
SOURce<hw>
.
1 | 2
Determines the baseband the connectors belong to, where:
SOURce1
= Baseband A/C and
SOURce2
= Baseband B/D
Parameters:
<Level>
float
Range:
0.3 to 2
Increment: 0.1
*RST:
1
Manual operation:
See
"Threshold Clock/Trigger Input"
[:SOURce<hw>]:INPut:TM:CLOCk:IMPedance
<Impedance>
[:SOURce<hw>]:INPut:TM:TRIGger:IMPedance
<Impedance>
Selects the input impedance for the external trigger/clock inputs.
Suffix:
SOURce<hw>
.
1 | 2
Determines the baseband the connectors belong to, where:
SOURce1
= Baseband A/C and
SOURce2
= Baseband B/D
SOURce subsystem