Remote control commands
R&S
®
SMM100A
724
User Manual 1179.1341.02 ─ 05
Example:
SOURce1:FREQuency:CW 2143E6
SCONfiguration:EXTernal:IQOutput1:RF:COUPling ON
SCONfiguration:EXTernal:IQOutput1:RF:FREQuency:OFFSet 20E6
// The resulting RF frequency of the external instrument is
// RF = 20E6 = 2.163 GHz
// Where both the RF frequency and the frequency offset
// are applied at the external instrument
Manual operation:
See
"(Delta) RF Frequency/ RF Level"
:SCONfiguration:EXTernal:FADer<ch>:RF:POWer
<Power>
:SCONfiguration:EXTernal:RF<ch>:RF:POWer
<Power>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:POWer
<Power>
In uncoupled mode, sets the RF level of the external instrument.
Parameters:
<Power>
float
Range:
-130 to 20
Increment: 0.01
*RST:
-30
Example:
SOURce1:POWer:LEVel:IMMediate:AMPLitude -30
SCONfiguration:EXTernal:IQOutput1:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput1:RF:POWer -20
// RF level of the external instrument is -20 dB
Manual operation:
See
"(Delta) RF Frequency/ RF Level"
:SCONfiguration:EXTernal:FADer<ch>:RF:POWer:OFFSet
<PowerOffset>
:SCONfiguration:EXTernal:RF<ch>:RF:POWer:OFFSet
<PowerOffset>
:SCONfiguration:EXTernal:IQOutput<ch>:RF:POWer:OFFSet
<PowerOffset>
In coupled mode, offsets the RF level of the external instrument with the selected delta
value.
Parameters:
<PowerOffset>
float
Range:
-100 to 100
Increment: 0.01
*RST:
0
Example:
SOURce1:POWer:LEVel:IMMediate:AMPLitude -30
SCONfiguration:EXTernal:IQOutput1:RF:COUPling ON
SCONfiguration:EXTernal:IQOutput1:RF:POWer:OFFSet -20
// RF level of the external instrument is -50 dB
// Where both the RF level and the level offset
// are applied at the external instrument
Manual operation:
See
"(Delta) RF Frequency/ RF Level"
SCONfiguration subsystem