Configuring the RF signal
R&S
®
SMM100A
402
User Manual 1179.1341.02 ─ 05
7.12.3.3
Fill with sensor
Access:
1. Select "RF" > "RF Level" > "User Correction".
2. Select "User Cor. Data > navigate to the file *.ucor > Select"
3. Select "Edit User Cor. Data > Fill With Sensor..."
This dialog contains parameters for filling a table automatically with sensor read-
ings.
To fill the table, select "Execute".
For information on power sensors and how to use them, see
calibrate the power level with an R&S NRP power sensor"
Settings:
Fill User Correction Data with Sensor
......................................................................... 402
Used SMM100A Settings For Measurement
...............................................................403
Fill User Correction Data with Sensor
●
"Sensor"
Displays connected sensors for selection.
●
"List To Fill"
Indicates the used list.
●
"S-Parameter"
Indicates whether S-Parameter correction in the R&S NRP power sensor is used.
●
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction data to improve
precision.
Improving level performance