Remote control commands
R&S
®
SMM100A
709
User Manual 1179.1341.02 ─ 05
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:NSRatio
<NSRatio>
Sets an upper limit for the relative noise content in fixed noise filter mode (
). This value determines the proportion of intrin-
sic noise in the measurement results.
Parameters:
<NSRatio>
float
Range:
0.001 to 1
Increment: 0.001
*RST:
0.01
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
<MTime>
Sets an upper limit for the settling time of the auto-averaging filter in the
NSRatio
mode and thus limits the length of the filter. The filter type is set with command
.
Parameters:
<MTime>
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:SONCe
Starts searching the optimum filter length for the current measurement conditions. You
can check the result with command
:SENS1:POW:FILT:LENG:USER?
in filter mode
USER
:SENSe<ch>[:POWer]:FILTer:TYPE
).
Example:
SENS1:FILT:TYPE USER
Selects user filter mode.
:SENS1:FILT:SONC
Activates the search for the optimum filter length.
:SENS1:FILT:LENG?
Returns the found optimum filter length.
Response: 128
Usage:
Event
SENSe, READ, INITiate and SLISt subsystems