General instrument functions
R&S
®
SMM100A
538
User Manual 1179.1341.02 ─ 05
Remote command:
[:SOURce]:INPut:USER<ch>:SIGNal
Execute Trigger
If "User 6 > Direction = Output" and "Signal = Manual Trigger", generates a short pulse
signal and outputs it at the User 6 connector.
The signal can serve as a common external trigger signal for triggering of several
R&S
SMM100A, see
Example "Triggering several R&S
Remote command:
:OUTPut:USER<ch>:TRIGger[:IMMediate]
Threshold User1-3 Input
Sets the high/low threshold in volts for the signal at the User1-3 connectors.
The input signal is defined with the parameter
. The same threshold applies for
all 3 connectors and any input signal.
Remote command:
[:SOURce]:INPut:USER:TRIGger:LEVel
[:SOURce]:INPut:USER:CLOCk:LEVel
Threshold User4-6 Input/Threshold Pulse Input
Sets the high/low threshold in volts for the signal at the User4-6 connectors.
The input signal is defined with the parameter
. The same threshold applies for
all 3 connectors and any input signal (e.g. external pulse modulation or trigger signal).
Remote command:
[:SOURce]:INPut:USER:PULM:LEVel
Impedance Clock/Trigger Input
Selects the input impedance for the external trigger/clock inputs.
Set the value to 1 k
Ω
/GND for high clock rates.
Remote command:
[:SOURce]:INPut:USER:TRIGger:IMPedance
[:SOURce]:INPut:USER:CLOCk:IMPedance
Clock Input Slope
Sets the polarity of the active slope of an externally applied clock signal.
Remote command:
[:SOURce]:INPut:USER:CLOCk:SLOPe
Trigger Input Slope
Sets the polarity of the active slope of an applied instrument trigger.
Remote command:
[:SOURce]:INPut:USER:TRIGger:SLOPe
Configuring local and global connectors