[:SOURce<hw>]:LIST:MODE
<Mode>
Sets the list mode.
The instrument processes the list according to the selected mode and trigger source.
See
LIST:TRIG:SOUR AUTO, SING or EXT
for the description of the trigger
source settings.
Parameters:
<Mode>
AUTO | STEP
AUTO
Each trigger event triggers a complete list cycle.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
*RST:
AUTO
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:POWer
<Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
Parameters:
<Power>
<Power#1>{, <Power#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also
.
Range:
depends on the installed options
Default unit: dBm
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
SOURce subsystem