Commissioning
PR 5500 Instrument Manual
EN-192 Sartorius
5.12.7 xBPI Calibration with Automatic Weight Detection
Requirements
-
The xBPI protocol has been selected (see Chapter 5.12.2).
-
The ‘xBPI scale’ weighing point has been selected.
-
The platform has been set up (see Chapter 5.12.4).
-
In menu [Weighing point A] ‘manual’ is set under
[Configuration]-[Weighing parameters]-[Confirming adjustment] (see page 187).
-
The communication between the device and platform is active.
Procedure
t
t
Select and confirm [Adjust with automatic
weight] using the cursor.
t
y
The calibration process is carried out without a
weight. The calibration status is displayed.
t
y
The value is specified automatically.
In this example, a weight of 10 kg is put onto the
scale.
t
y
The following window is displayed after applying
the weight.
t
y
The deviation is displayed in the last line with
increased resolution (10-fold).
t
t
Press the [Accept] soft key.