Device Installation
PR 5500 Instrument Manual
EN-82 Sartorius
Connection example for PR 5500/13:
Relay control (current output)
Connection example for PR 5500/13:
Voltage output
CH4B
CH4C
CH 2
CH 3
CH 1
*
* inductive load/induktive Last
supression diode/Freilaufdiode
-
+
PR 1624
24 V DC 0.5 A
grounding terminal or
screen clamping rail/
Erdungsklemme oder
Schirmklemmschiene
*
* 2.2 k¥/1 k¥ with 24 V/12 V
2,2 k¥/1 k¥ mit 24 V/12 V
-
+
PR 1624
24 V DC 0.5 A
-
+
CH4B
CH4C
CH 2
CH 3
CH 1
grounding terminal or
screen clamping rail/
Erdungsklemme oder
Schirmklemmschiene
The relay switches when the output is active (true). To
protect the output circuit, relays must be equipped
with free-wheel diodes.
When the output is active (true), the output voltage
goes from 24 V/12 V to 0 V.
The load resistance must be 2.2 k¥/1 k¥.