Commissioning
PR 5500 Instrument Manual
EN-218 Sartorius
5.15 Calibrating Digital Precision Compression Load Cell ‘Pendeo® Process’
5.15.1 General Information
The digital load cells have been calibrated at the factory based on the acceleration of gravity at Hamburg,
Germany: 9.81379 m/s². The calibration data in the load cells are invariable. The calibration data for the gravity
acceleration at the place of installation can be adapted only in the instrument and protected against overwriting
(see Chapter 5.1.5). With applications for use in legal metrology, the legal requirements and the conditions given
on the test/approval certificate must be taken into account when selecting the settings.
5.15.2 Parameter Input for Serial Interface
t
t
Go to [System setup]-[Weighing point] [Weighing
point A], select and confirm ‘Pendeo Process’.
t
y
The following window appears.
t
t
Select and confirm the line ‘Interface’.
t
y
The following window appears.
t
t
Select and confirm the RS-485 interface (in this
case: Option-1 RS-485-A).
t
y
The previous window appears.
t
t
Press the [Save] soft key to save the changes.