PR 5500 Instrument Manual
Commissioning
Sartorius EN-123
5.6.6 Resetting Network
Note
Pressing the reset key during a long time (>5 sec.) triggers a cold start.
-
The settings will not be changed.
-
The database will be emptied.
-
Current process steps are deleted.
-
Application must be restarted.
-
The database must be restored.
Pressing the reset switch during a long time (>5 sec.) resets the network settings to default/factory settings.
That means:
-
LED will be switched on a second time.
-
‚DHCP‘ is activated.
-
Web server is enabled.
-
VNC access is enabled.
-
‚Host name’ is initialized e.g. PR5500-6B6A5E ( device type-MAC-ID).
Example of MAC-ID: 00-90-6C-
6B-6A-5E
This ensures that a valid address for identification of the instrument in the network can be assigned to the
device, see Chapter 5.19.4.
Note
The last 3 bytes of the MAC ID are displayed. A label with the complete MAC ID is fitted
to the outside of the device.
A device set to DHCP
(default/factory setting) and connected to an IT network
(company network) with a DHCP server does not require further actions except for
a
2…3-minute waiting time
. Subsequently, a network connection is established
automatically (device <-> workstation/PC).