Error Messages
Sartorius EN-381
11.6.3 Determining the Dead Load
The dead load entered in mV/V plus maximum capacity
in mV/V is higher than 3 mV/V (= 36 mV).
This message appears when the scale is not stable.
Remedy
t
t
Check the mechanical function of the scale.
t
t
Adapt the filter setting; reduce the resolution.
t
t
Adapt the stability conditions.
This message appears if the measurement signal is
negative (load cells connected with wrong polarity or
defective) when determining the dead load with [by
load].
Cause
Load cell connected with wrong polarity, or defective,
or mechanical problem of the scale.
This message appears if the dead load entered in mV/V
is higher than 5 mV/V.