Commissioning
PR 5500 Instrument Manual
EN-118 Sartorius
5.1.4 Cold Start
5.1.4.1 General Information
There are several options executing a cold start:
-
After power-on (see Chapter 5.1.4.2)
-
With STOP+EXIT keys (see Chapter 5.1.4.3)
-
With reset key (see Chapter 5.1.4.4)
5.1.4.2 Menu-driven Power Off
The menu-driven power off (see Chapter 6.1.9) does
not
save the entire content of the SD-RAM into a Nand
flash memory. However during the cold start the last saved database is automatically loaded into the device.
5.1.4.3 With STOP+EXIT Keys
During the cold start with STOP+EXIT keys (see Chapter 5.19.3) the database will be emptied and current process
steps are deleted. The device automatically searches for an existing database on the SD card and ask to load it.
Note
If ‚disabled‘ is selected under [System setup]-[Operating parameters]-[Coldstart with
STOP+EXIT], the STOP key must be pressed while the power cord is being connected to
execute a cold start.
5.1.4.4 With Reset Key
The device re-started (cold start) by a short-time (<1 sec.) actuating of the reset key (see Chapter 3.3.1).
-
LED will flash once.
-
No settings will be changed.
-
The database will be emptied.
-
Current process steps are deleted.
-
Application must be restarted.
-
The database must be restored.
-
No network settings will be changed.