PR 5500 Instrument Manual
Commissioning
Sartorius EN-159
5.11.5 Determining the Maximum Capacity [Max]
The maximum capacity (Max) determines the maximum weight without dead load of the weight to be measured
and the displayed number of digits behind the decimal point. Normally, Max is less than the load cell capacity
(nominal capacity x number of load cells).
Permissible values for the maximum capacity are:
Max weight value from 0.00010 to 999999 in t, kg, g,
mg, lb or oz.
Max weight value must be an integer multiple of the
scale interval (1 d). It may have up to 6 digits and is
entered as a numeric value with or without a decimal
point.
t
t
Enter the maximum capacity with decimal places
(in this example: 3000.0).
t
t
Double click key to select the weight unit.
t
t
Confirm the entries.
t
y
The confirmation is displayed with “setting Max...”.
t
t
Press the [Save] soft key to save the calibration.
t
y
The reference value is calculated for the
electronics test.
Note
If linearization is active (see Chapter 5.11.12): After selection of the line "Max at" the
following tip is displayed:
"cannot be changed here
while linearization is active"
Only deletion of linearization points deactivates linearization!