PR 5500 Instrument Manual
Commissioning
Sartorius EN-125
5.9 User Management
5.9.1 General Information
By means of the user management, the access to various menu levels of the instrument can be limited, if the
necessity arises. This includes e.g. system set-up, system maintenance, access via the website and the various
levels of the application (application-specific).
Note
The user management can be activated optionally. However, it is not required for normal
operation.
During system installation and set-up, we recommend leaving the user management de-activated. This facilitates
the access to the various menus and to the website.
Always activate the user management, if
-
several persons work with the unit or have access to it.
-
the target is to prevent unauthorized persons from making changes to the unit or from influencing the
processes controlled by the unit.
5.9.2 Activating User Management
User management is
not
activated by default.
t
t
Select and confirm the [System setup]-[User
management] menu item.
t
y
A prompt window appears.
t
t
Press the corresponding soft key.