PR 5500 Instrument Manual
Commissioning
Sartorius EN-157
5.11.4 Selecting the Calibration Mode
Note
The menu item ‘Modify’ is only used for small changes (e.g., changing the dead load/preload,
changing the mV/V values for dead load/preload and/or Max, changing the scale interval).
Otherwise selected ‘New’.
Go to [System setup]-[Weighing point]-[Weighing
point A]-[Calib] and use the soft keys to choose
between [New] and [Modify].
5.11.4.1 New Calibration
t
t
Select and confirm [System setup]-[Weighing
point]-[Weighing point A]-[Calib].
t
t
Press the [New] soft key.
When you press [New] , the default settings are
restored and then calibration is started.
t
y
You are prompted to confirm.
t
t
Press [Continue] for the default settings or
[Cancel] to cancel the selection.