Device Installation
PR 5500 Instrument Manual
EN-104 Sartorius
4.4.12 PR 1721/57 EtherNet-IP
4.4.12.1 Specifications
The fieldbus card is equipped with a standard RJ-45 socket for the Ethernet connection and contains a powerful
TCP/IP and EtherNet IP connecting circuitry with transfer rates of 10 and 100 Mbps.
The card is inserted in the Option-1/FB slot, see Chapter 13.6.
Internal connection
Pin strip, 50-pin
External connection
RJ-45 connection socket in the module cover
Transmission rate
10 Mbps and 100 Mbps
Autodetection (10/100, HalfDX/FullDX)
Connection mode
Network
Protocol
EtherNet IP
Configuration
EDS file
'sag_5500 ethernetip.eds'
Cables
Twisted pairs, screened e.g. patch cable CAT5 Autolink (straight or crossover)
Cable impedance
150 ¥
Cable length to HUB
Max. 115 m
Potential isolation
Yes
Dimensions (LxWxH)
55x50x22 mm
Weight
Approx. 33 g
Certificates
EtherNet IP Specification ODVA File No. 10286
Test Date: 06.09.2005
Vendor ID 90
See also: www.odva.org
Tested according to: CE, UL & cUL
Note
The IP address and the Subnet mask are set via the Fieldbus parameters (see also
Chapter 5.19.6).
The EDS file is stored on the CD supplied with the device (Fieldbus directory of the respective
device). The current file is also available for download via the Internet:
http://www.sartorius-intec.com/en/products/software-tools/
More -> Fieldbus-files