PR 5500 Instrument Manual
Commissioning
Sartorius EN-197
t
y
The second linearization point to be calibrated is
displayed.
t
t
Place the corresponding weight (in this case, 5 kg)
on the scale.
t
y
The deviation is displayed in the last line with
increased resolution (10-fold).
t
t
Press the [Accept] soft key.
t
y
The third linearization point to be calibrated is
displayed.
t
t
Place the corresponding weight (in this case,
10 kg) on the scale.
t
y
The deviation is displayed in the last line with
increased resolution (10-fold).
t
t
Press the [Accept] soft key.
t
y
The last linearization point to be calibrated is
displayed.
t
t
Das entsprechende Gewicht (hier 30 kg) auflegen.
t
y
Die Abweichung wird in der letzten Zeile mit
10-fach Auflösung dargestellt.
t
t
Softkey [Annehmen] drücken.
t
t
Taste ESC/EXIT drücken, um zum vorherigen
Fenster zurückzukehren.
Summary of Contents for Sartorius Intec Maxxis 4 PR 5500