Commissioning
PR 5500 Instrument Manual
EN-224 Sartorius
5.15.7 Calibrating Load Cells
Access the Pendeo setup menu as follows: [System setup]-[Weighing point] [Weighing point A]-[Calib].
Example
Max. load cell capacity: E
max
= 5 t
Number of load cells: 4
Number of vessel feet: 4
Max: 20.000 t
Scale interval: 0.020 t
Dead load: empty weight
Calibration weight: 5.000 t
Procedure
t
t
Select and confirm the menu item ‘Calibrate the
scale’.
t
t
Press the [New] soft key.
Note
The menu item ‘Modify’ is
only used for small changes
(e.g., changing the dead load/
preload).
Otherwise selected ‘New’.
t
y
A prompt window appears.
t
t
Press the [Yes] soft key.