Commissioning
PR 5500 Instrument Manual
EN-240 Sartorius
5.18 Setting the SMA Scale
5.18.1 General Information
The PR 5500 can communicate with a scale via the SMA protocol. The device can be connected via serial
interface (PR 5500/04; see Chapter 4.4.1). Communication is serial.
5.18.2 Parameter for Serial Interface
t
t
Go to [System setup]-[Weighing point]-[Weighing
point A], select and confirm ‘SMA scale’.
t
y
A selection window appears.
Unavailable parameters are grayed out.
t
t
Select and confirm the individual settings.
[Interface]
serial interfaces
Selection:
not assigned, Built-in RS-232, Option-x RS-485-A,
Option-x RS-485-B
[
Baudrate
]
Modulation rate for data transmission.
Note
The selected value must match the value of the connected device.
Selection:
300, 600, 1200, 2400, 4800, <9600>, 19200 baud
[
Parity
]
Parity check for detecting errors during data transmission.
Selection:
no parity, odd parity, even parity
[Default]
Settings are reset to factory settings.
[Save]
The settings are saved.