PR 5500 Instrument Manual
Error Messages
Sartorius EN-379
11.6 Error Messages during Calibration
11.6.1 Determining the Maximum Capacity (MAX)
The maximum capacity of the scale can be increased
retroactively.
This message appears when the measurement signal
for the given maximum capacity would exceed the
permissible input voltage.
The maximum capacity of the scale can be increased
retroactively. When the capacity is reduced, however,
a message appears when the new maximum capacity
falls below the calibration weight ("calibrated at").
This message appears if the selected resolution is too
low, e.g. 5 kg.
This message appears when the selected resolution
is so high that less than 0.8 internal counts per scale
interval (d) are available.