PR 5500 Instrument Manual
Commissioning
Sartorius EN-179
5.11.16.1 Operation in Legal Metrology
Note
The following settings are available; but the Maxxis 4 is
not
legal for trade!
Go to [System setup]-[Weighing point]-[Weighing point A]-[Parameters]-[W&M] and choose between:
[none] and the legal metrology modes [OIML], [NTEP], or [NSC].
[none]
[OIML]
[NTEP]
[NSC]
Gross weight display B
B
G
G
Recommended min.
measurement signals
0.2 mV/V @ 30,000 d 0.2 mV/V @ 3,000 e 0.2 mV/V @ 3,000 e 0.2 mV/V @ 3,000 e
0.4 mV/V @ 60,000 d 0.4 mV/V @ 6,000 e 0.4 mV/V @ 6,000 e 0.4 mV/V @ 6,000 e
If operation for legal metrology is switched on, the parameter settings (zero tracking etc.) must be selected
accordingly. The device does not carry out a check of this.
A unique check number is created every time a calibration or changed parameters are saved. This can be
viewed under [System information]-[Show calibration check numbers] (see Chapter 5.20.7) and noted. See also
Chapter 5.11.1.