PR 5500 Instrument Manual
Commissioning
Sartorius EN-207
t
y
The new assignment will be displayed.
t
t
Check the corner load (dead load), see
Chapter 5.14.10.1.
t
t
Press the ESC/EXIT key to return to the previous
window.
5.14.7 Calibrating Load Cells
Access the Pendeo setup menu as follows: [System setup]-[Weighing point]-[Weighing point A]-[Calib].
Example
Max. load cell capacity: E
max
= 50 t
Number of load cells: 4
Max: 200.000 t
Scale interval: 0.020 t
Dead load: empty weight
Calibration weight: 11.000 t
Procedure
t
t
Select and confirm the menu item ‘Calibrate the
scale’.
t
t
Press the [New] soft key.
Note
The menu item ‘Modify’ is
only used for small changes
(e.g., changing the dead load/
preload).
Otherwise selected ‘New’.