PR 5500 Instrument Manual
Commissioning
Sartorius EN-199
5.13 Setting the SBI Scale
5.13.1 General Information
The PR 5500 can communicate with scales (e.g., Combics1) or weighing modules via the SBI protocol. The scale
can be connected via serial interface (PR 5500/04; see Chapter 4.4.1). Communication is serial. There are no
variables or digital I/Os for external scales in the SPM of the PR 5500.
5.13.2 Parameter for Serial Interface
t
t
Go to [System setup]-[Weighing points] [Weighing point A], select and confirm ‘SBI scale’.
t
y
A selection window appears.
Unavailable interfaces are grayed out.
t
t
Select and confirm the individual settings.
[Interface]
serial interfaces
Selection:
not assigned, Built-in RS-232, Built-in RS-485,
Option-x RS-485-A, Option-x RS-485-B
[
Baudrate
]
Modulation rate for data transmission.
Note
The selected value must match the value of the connected device (e.g. Combics1).
Selection:
300, 600, 1200, 2400, 4800, <9600>, 19200 baud
[
Data bits
]
Groups of data bits.
Selection:
7 data bits, <8 data bits>
[
Parity
]
Parity check for detecting errors during data transmission.
Selection:
no parity, odd parity, even parity
[Stop bits]
Units for transmission protocols.
Selection:
1 stop bit, <2 stop bits>
[Default]
Settings are reset to factory settings.
[Save]
The settings are saved.