PR 5500 Instrument Manual
Device Installation
Sartorius EN-105
4.4.12.2 LEDs
8
1
MS
NS
EtherNet/IP
Link
Identification
Description
NS
Network status LED
MS
Module status LED
Link
Link/Activity LED
Network status (NS)
LED status
Description
Comments
Off
Module is offline
-
No power
-
No IP address
Constant green
Module is online
One or more connections exist (chip class
1 or 3)
Flashing 1 Hz green
Module is online
No connections
Constant red
Critical connection error
Duplicate IP address
Flashing 1 Hz red
Connection time exceeded
For one or more connections (chip class 1
or 3)
Module status (MS)
Status
Description
Comments
Off
Module is not initialized.
-
No power
-
Module is in the SETUP or NW_INIT
status.
Constant green
Module is initialized.
Normal operation
Flashing 1 Hz green
Error after test
Error occurred after test.
Flashing 2 Hz green
This is used for node identification in the
network.
Constant red
Exception error
-
Module is in the EXCEPTION status.
-
The exception error monitoring system
closes all open connections to the
module.
Flashing 1 Hz red
Configuration error
Expected identification deviates from the
available identification.
Flashing 2 Hz red
IP address error
IP address was not defined.
Flashing 3 Hz red
Device name error
Device name was not defined.
Flashing 4 Hz red
Internal error
Module has caused an unrecoverable
internal error.