Commissioning
PR 5500 Instrument Manual
EN-236 Sartorius
5.17 Mettler-Scale
5.17.1 General Information
PR 5500 can communicate with the Mettler-Scale via the MT-SICS protocol. The scale can be connected via
serial interface. Communication is serial.
The determination and transmission of weight depends strongly on the scale. Weight values up to seven digits
plus pre/- sign can be displayed. The maximum capacity (Max) is already set in the Mettler scale and
cannot be changed via the MT-SICS protocol.
5.17.2 Parameter Input in Mettler-Scale Menu
Example: scale type XS6002SDR
Select the menu item ‘Peripherals’ in system menu and make the following settings:
Printer
Off
Host
RS232 built-in
Define
Baudrate
9600
Bit/Parity
7/Odd
Stop Bits
1 Stop Bit
Handshake
Xon/Xoff
End of line
<CR> <LF>
Char Set
IBM/DOS
Continuous mode
Off
Secondary Display
Off
Bar Code
Off
Ext. Keyboard
Off
5.17.3 Parameter Input for Serial Interface
t
t
Go to [System setup]-[Weighing point]-[Weighing
point A], select and confirm ‘Mettler-Scale’).
t
t
The following window appears.
t
t
Select and confirm the line ‘Interface’.