Device Installation
PR 5500 Instrument Manual
EN-102 Sartorius
4.4.11.2 LEDs
8
1
MS
NS
ProfiNet I/O
Link
Identification
Description
NS
Network status LED
MS
Module status LED
Link
Link/Activity LED
Network status (NS)
LED status
Description
Comments
Off
Module is offline
-
No power
-
No connection to the I/O controller
Constant green
Module is online (RUN)
-
There is a connection to the
I/O controller
-
I/O Controller is operational (RUN
status)
Flashing 1 Hz green
Module is online (STOP)
-
There is a connection to the
I/O controller
-
I/O Controller is not operational (STOP
status)
Module status (MS)
Status
Description
Comments
Off
Module is not initialized
-
No power
-
Module is in the SETUP or NW_INIT
status
Constant green
Module is initialized
Normal operation
Flashing 1 Hz green
Error after test
Error occurred after test.
Flashing 2 Hz green
This is used for node identification in the
network.
Constant red
Exception error
-
Module is in the EXCEPTION status
-
The exception error monitoring system
closes all open connections to the
module.
Flashing 1 Hz red
Configuration error
Expected identification deviates from the
available identification.
Flashing 2 Hz red
IP address error
IP address was not defined.
Flashing 3 Hz red
Device name error
Device name was not defined.
Flashing 4 Hz red
Internal error
Module has caused an unrecoverable
internal error.